Tungsten Eyeshield Backscatter Data |
|
The doses are normalized to dmax without the eye shield (Diagram 1) using a 10 x 10 cone. When 100% is delivered to dmax using 6 MeV with the shield, you get 108% to the undersurface of the eyelid (Point 1) and 3.4% to the lens (Point 2) (See table).
Unreferenced data on this product is preliminary findings of Radiation Products Design, Inc. and is not to be used as a technical reference. *XV-2 Film placed under/below tungsten eye shield at 3 mm depth (anterior surface of lens) **TLD Microcubes placed under simulated eye lid using tungsten eye shields. Reference: Evaluation of Eye Shields made of Tungsten and Aluminum in High-Energy Electron Beam - Randi D. Weaver B.S. Fairview - University Med. Ctr. PO Box 494, 420 Delaware St. SE, Mpls., MN 55455 Int. J. Radiation Oncology Biol. Phys, Vol. 41 Nal, pp 233-237-1998
|
|||||||||||||||||||||||||||||||||||||||
![]() |