Isolator
- Sealed glove box design with sealed interchange chamber
- Unit shall maintain ISO Class 5 (Federal Standard 209E Class 100) within workzone and interchange chamber
- Interchange chamber may be configured on the right or left side
- Negative pressure, with sufficient interchange chamber pressure differential to maintain sterility in glove box
- Interchange door is interlocked to prevent inadvertent opening when outer door is open
- Two removable work trays with cove corners for main and interchange chambers
- HEPA filter for supply air protected by non flammable perforated metal diffuser
- HEPA filter seal is spring loaded Neoprene
- Partial re-circulating airflow with minimal room air exhaust
- Externally mounted fluorescent lighting
- Hospital grade duplex outlet on back wall with drip proof cover
- Hinged front face, ergonomically slanted at 10° from vertical
- 16 GA, Type 304 stainless steel liner with reinforced U channels
- Top/Front filter removal access without disassembly of control panel
- Two 8 inch diameter glove ports
- Two circular nitrile rubber gloves
- Two circular nitrile rubber sleeves
- Three position IV bar with six hooks
- Key locks included on external front face
Shielding/Base
- Viewing window: 1/2 in. of 4.8 g/cm3 leaded glass, 25 in. wide x 7 1/2 in. height
- 1/8 in. lead shielding on base, back and sides (optional 1/4 in.) 1/8 in. lead shielding on hinged front panel excluding glove ports and viewing window
- 1/4 in. lead body shield in center of front panel
- Two 8 in. diameter penetrations in base for sharps container and flush mount dose calibrator chamber
- 1/8 in. lead shielding for sharps container
- Articulating arm for calibrator readout mounted on base
- Exterior aluminum cosmetic panels painted epoxy enamel
- Heavy duty base support-adjustable height with 10 in. range
- Height adjustment to center of glove port-42 1/2 in. to 52 1/2 in.
|
|
Electrical/Control Panel
- Negative pressure design
- Exhaust interlock to prevent positive pressure workzone and interchange
- Centrally located control panel for adjustable airflow and lights
- Airflow adjustment via solid state motor voltage regulator
- Fuses located in control panel
- Two minihelic gauges
- Meets UL requirements for U.S. and Canada for electrical safety
- UL/UL-C Listed 115 V, AC 60 Hz, 6 Amps (220V/50 Hz option)
- 12 ft. Hospital Grade Power Cord
- Exhaust rate 65 CFM (110CMH) @ 0.1in. w.g. (external exhaust option only)
- Heat Rejected, 795 BTU per hour
Dimensions
- Overall Dimensions: 51 1/2 in. wide x 31 in. depth x 80 1/2 in.- 90 1/2 in. height, depending upon adjustable base setting
- Workzone Dimensions: Main Cell 35 1/4 in. wide x 23 1/4 in. depth x 27 3/8 in. height
- Interchange chamber Dimensions: 14 1/8 in. wide x 24 in. depth x 27 3/8 in. height
- Interior sliding door: 18 in. wide x 8 1/2 in. height
- Weight: 1/8 in. 1,200 pounds, 1/4 in. option 1,600 pounds
Installation
- Unit is designed for customer installation
- Disassembled unit will fit through standard 36 in doorway
- Base adjustment must be set before placing shielded glove box
Options
- Thimble kit for external exhaust
- Additional lead shielding (1/4 in. total) on sides, floor, and sharps shield
|