CAP-Cell® Isolator

Isolator

  • Sealed glove box design with sealed interchange chamber
  • Unit shall maintain ISO Class 5 (Federal Standard 209E Class 100) within workzone and interchange chamber
  • Interchange chamber may be configured on the right or left side
  • Negative pressure, with sufficient interchange chamber pressure differential to maintain sterility in glove box
  • Interchange door is interlocked to prevent inadvertent opening when outer door is open
  • Two removable work trays with cove corners for main and interchange chambers
  • HEPA filter for supply air protected by non flammable perforated metal diffuser
  • HEPA filter seal is spring loaded Neoprene
  • Partial re-circulating airflow with minimal room air exhaust
  • Externally mounted fluorescent lighting
  • Hospital grade duplex outlet on back wall with drip proof cover
  • Hinged front face, ergonomically slanted at 10° from vertical
  • 16 GA, Type 304 stainless steel liner with reinforced U channels
  • Top/Front filter removal access without disassembly of control panel
  • Two 8 inch diameter glove ports
  • Two circular nitrile rubber gloves
  • Two circular nitrile rubber sleeves
  • Three position IV bar with six hooks
  • Key locks included on external front face

Shielding/Base

  • Viewing window: 1/2 in. of 4.8 g/cm3 leaded glass, 25 in. wide x 7 1/2 in. height
  • 1/8 in. lead shielding on base, back and sides (optional 1/4 in.) �� 1/8 in. lead shielding on hinged front panel excluding glove ports and viewing window
  • 1/4 in. lead body shield in center of front panel
  • Two 8 in. diameter penetrations in base for sharps container and flush mount dose calibrator chamber
  • 1/8 in. lead shielding for sharps container
  • Articulating arm for calibrator readout mounted on base
  • Exterior aluminum cosmetic panels painted epoxy enamel
  • Heavy duty base support-adjustable height with 10 in. range
  • Height adjustment to center of glove port-42 1/2 in. to 52 1/2 in.
 

Electrical/Control Panel

  • Negative pressure design
  • Exhaust interlock to prevent positive pressure workzone and interchange
  • Centrally located control panel for adjustable airflow and lights
  • Airflow adjustment via solid state motor voltage regulator
  • Fuses located in control panel
  • Two minihelic gauges
  • Meets UL requirements for U.S. and Canada for electrical safety
  • UL/UL-C Listed 115 V, AC 60 Hz, 6 Amps (220V/50 Hz option)
  • 12 ft. Hospital Grade Power Cord
  • Exhaust rate 65 CFM (110CMH) @ 0.1in. w.g. (external exhaust option only)
  • Heat Rejected, 795 BTU per hour

Dimensions

  • Overall Dimensions: 51 1/2 in. wide x 31 in. depth x 80 1/2 in.- 90 1/2 in. height, depending upon adjustable base setting
  • Workzone Dimensions: Main Cell 35 1/4 in. wide x 23 1/4 in. depth x 27 3/8 in. height
  • Interchange chamber Dimensions: 14 1/8 in. wide x 24 in. depth x 27 3/8 in. height
  • Interior sliding door: 18 in. wide x 8 1/2 in. height
  • Weight: 1/8 in. 1,200 pounds, 1/4 in. option 1,600 pounds

Installation

  • Unit is designed for customer installation
  • Disassembled unit will fit through standard 36 in doorway
  • Base adjustment must be set before placing shielded glove box

Options

  • Thimble kit for external exhaust
  • Additional lead shielding (1/4 in. total) on sides, floor, and sharps shield

 

    