PET-CT Phantom™

The PET-CT Phantom™ includes internal structures (three rods and six spheres) which, when imaged with both modalities, can demonstrate how accurately the two image sets are aligned.

A single sample of radioactive water is attenuated by water, bone and CT contrast material (as well as air only) to determine how accurately the CT-based PET attenuation correction works.

The Phantom is used for:

  • Acceptance testing of PET/CT and SPECT/CT systems
  • Routine quality evaluation of PET/CT and SPECT/CT systems
  • Evaluation of new image fusion software
  • Evaluation of new attenuation correction algorithms
  • Aluminum tubes evaluate registration
  • Outer 2" OD micro cylinder compares attenuation region to non attenuation region
  • 6" ring studies contrast solution
  • Research

043-771 PET-CT Phantom

 

PET/CT Phantom

Specifications
Main Cylinder Interior length of phantom 180 mm; Fillable spheres (5) inner diameter 10 mm, 13 mm, 17 mm, 22 mm, and 28 mm.
Distance from sphere plane to inside wall 70 mm
Volume of empty cylinder 9.7 liters Main Cylindrical insert dimension Outside diameter 51 mm; Length 180 mm
Top Cylinder Cylinder outside diameter ~5.1 cm; Cylinder inside diameter ~4. cm; Cylinder inside height ~8.2 cm; Cylinder outside height ~12.0 cm; Volume of empty cylinder 408 cm2
Three Aluminum Tubes One 5 inch long ~1.7 cc; Two 7 inch (ea) ~ 2.5 cc Stepped Bone Ring Pre-filled with liquid bone composition, not to be opened. The volumes for the bone ring are: Outer volume, 15.6 cu inch: ~256 cc; Inner volume 6.7 cu inch, ~110 cc