Xenon Air Monitor |
|
Range 4 decades: 0.1 mpc (10-6 µCi/ml) to 1000 mpc (10-2 µCi/ml) Sensitivity 0.1 mpc Detection: Gas chamber with 2” Pancake GM detector Shielding 3/8” thick around chamber and detector (7” h x 4.5” dia.) Flow Rate Regulated - normally set at 10 lliters/ minute Prefilter Used to remove particulates |
Case Aluminum, 11” x 11” x 11” Display 6 digit LED Air Pump Vane type oil-less; muffler included (6” x 6” x 6”) Alarm Audible 200 Hz beeper; Visual red flasher set at any point on scale Power 115V, 60 Hz Weight Electronics 10 lb; Shielded Detector 22 lb; Pump 7 lb |