JRT Associates   Nuclear Medicine
Radiation Oncology
Emergency Preparedness
PET Imaging
Nuclear Cardiology
Medical Physics
Home Products Contest Central Newsletters JRT Calibrations Contact JRT

ASM990 and Model 190 Survey Meter Detectors

Pancake GM Detector

Pancake GM Detector

Operating Voltage 900 + 50 Vdc
Plateau Length 100 Volts minimum
Plateau Slope 0.1 % per V maximum
Dead Time 50 μs maximum
Temperature Range -30 °C to 75 °C (-22 °F to
167 °F)
Mica Window Thickness 1.4 to 2.0 mg/cm2
Mica Window Size 4.45 cm dia. (1.75”)
Active Area 15.5 cm2 (2.4 in2)
Beta Efficiencies 137Cs = 22%, 60Co = 16%, 90Sr 90Y = 32%, 99Tc = 15%, 14C = 6%
Alpha Sensitivity 3 MeV at window
Connector BNC
Size (9.75” L x 2.68” W x 2.75” H)
Weight 0.23 kg (0.5 lb)

489-110D Pancake Probe $489.00

 


 
End Window GM Detector

End Window GM Detector

Radiation Detected Alpha > 3 MeV, Beta > 45 keV, Gamma > 6 keV
Window/Wall Material Mica
Window/Wall Thickness 1.4 to 2 mg/cm2
Window Area 6.8 cm2
Dead Time 200 µs nominal
Saturation > 5R/h
Efficiency C-114 10%; 147-Pm 20%; 99-Tc 35%; 36 Cl 35%; 90-Sr/ 90-Y 45%; 239-Th 50%
Size/Weight 4.5” x 1.4”; 5.5 oz

489-35 Thin End Window Probe $990.00

 


 
Scintillation Detector

Scintillation Detector

Type NaI (Tl) Sodium Iodide 1 x 1, scintillator optically coupled to PMT
Radiation Detected Gamma and x-ray > 60 keV
Typical Background (cpm) 1750
Nominal Sensitivity 160,000 cpm/mR/hr 137Cs
Wall Material 0.04 in Al, 1 mm thick
Window 108 mg/cm2 Al
Sensitive Area 5 cm2
Crystal Dimensions 1 x 1 in.
Probe Diameter 2 in.
Probe Length 8.75 in
Cable Length 48 in.
Operating Voltage 900 V
Calibration 137Cs 2 pts/scale to 10 mR/hr
Efficiency 137 Cs 6%; 57Co 9%; 133Ba 6%;
60Co 2%
Operating Temperature - 40° to + 120°F
- 40° to + 50°C. Max. temp increase of 20°F/hr
Weight 1.5 lb (0.68 kg)

489-50 NaI Scintillation Probe, 1” x 1” $1,279.00

 


 
Energy Compensated GM Detector

Energy Compensated GM Detector

Radiation detected Beta above 200 keV, Gamma above 12 keV
Operating voltage 900 V
Plateau length 100 V (min)
Plateau slope 0.1%/V (max)
Temperature range - 40° to + 167°F
(- 40° to + 75°C)
Relative humidity 0 to 95%, noncondensing
Wall thickness 30 mg/cm2 (tube)
Wall material Stainless steel
Maximum background 15 CPM
Nominal sensitivity 720 CPM/mR/hr
Maximum exposure rate 1 R/hr

90-12 Energy Compensated GM Detector $888.00