RO-20 Ion Chamber Survey Meter |
| ENERGY RESPONSE Photon Response Reference to 137Cs measured through the bottom with the slide closed, the energy response is: ± 30% from 8 keV to 1.3 MeV with the open slide facing the source. ± 15% from 33 keV to 1.3 MeV with the closed slide facing the source. ± 15% from 55 keV to 1.3 MeV through the side of the instrument. Beta Response Uranium Slab: 30% of true mrad/h field behind 7 mg/cm2 window with RO-20 resting on slab, slide open. 90Sr90Y: Approximately 93% of true mrad/h field at 30 cm with slide open. Fast Neutron Response (PuBe) Reads approximately 8% in mR/h of true neutron field in mrem/h. RADIOLOGICAL Radiation Detected Beta, gamma, and x-ray. Ranges Five linear ranges: 0-5, 0-50, 0-500 mR/h and 0-5, 0-50 R/h. Meter Scale length, approx. 7.6 cm (3”), 2% accuracy. Linear markings from 0 to 5 in 50 minor increments. The meter is back-lit. Response Time 90% of final reading within 5 seconds; see options below for faster response. Linearity Within ± 5% of full scale. Battery Dependence Reading is independent of battery voltage when the battery check indication is in the green arc. External Controls Range switch, including Off, Zero, and Battery checking positions. Zero knob used to set meter to zero when Zero position of range switch is selected or when in no significant radiation field. Light switch, for meter light. Internal Controls Five calibration controls, one for each range. |
Batteries Main power: Five "C" cells. |