- Complete remote radiation information management for Model
451P and 451B Ion Chamber Survey
Meters
- Color-coded data provides quick identification for radiation levels and alarm acknowledgment status
- Real time data logging in protected Excel worksheet
- Virtual instrument display with audible and
visual alarm indication
- Remote user definable alarm settings
- Downloads 451’s internal data log into Excel worksheet
- Provides remote control and instrument configuration
- Complete on-line help speeds learning
- Compatible with Windows® 95, 98, ME®, NT® 4.0, 2000, XP® and Excel 97, 2000
- Package includes manual, diskette set, and
25 ft RS-232 cable
The 451 Assistant for Excel is an Excel Add-in, a program that adds optional commands and
features to Microsoft® Excel. The 451 Assistant provides remote control for many of the 451
functions, real-time data logging with user defined alarm parameters and a real time virtual
instrument display. The 451 Assistant automatically loads with Excel once it is installed.
The 451 Assistant data logging function automatically records real time measured data in an
active Excel worksheet. Each measurement in the data log is automatically placed in the active
worksheet cell, then the active cell moves down to the next row of cells where it will place the
next set of measured data. Each recorded measurement is time and date stamped.
The
451 Assistant provides user configurable audible and visual alarms for the real time logged
data. The picture and data shows radiation levels above the preset alarm in red, the blue data
represents the radiation level below the alarm that has not been acknowledged. The black data
shows when the alarm was acknowledged. See Control Options for more details.
The 451 Assistant for Excel allows the user to configure the 451’s internal data logging and
alarm settings. It may also be used to download the 451’s internal data log and saves it in an
Excel worksheet where it can be easily graphed for quick trend analysis.
The 451 Assistant provides automatic timed integrated dose measurement over a user defined
integration period or user controlled integrated dose measurement. After the integration timer
has expired or the user has stopped integration, the 451 Assistant retrieves the integrated dose
from the 451 and inserts the integrated dose in the active worksheet cell and the average dose
rate is placed in the next cell to the right.
The Model 451EXL information management software program is ideal for the facility
Radiation Safety Officer or anyone responsible for maintaining a permanent record of spills
and accidents for adherence to state and NRC requirements. The remote control of the 451
series of Ion Chambers may be accessed by any computer on your facility network providing
remote inform-ation management.
451EXL 451 Assistant for Excel $536.00
|
|

Click image to enlarge
Controls The 451 Assistant menu and toolbar provide an interface for the user
to remotely control the 451, configure the 451, ownload the 451’s
internal data log and start / stop real time data logging and integrated
dose measurements.
Connect / Disconnect Connects or disconnects the 451 to the
computer’s communication port.
Start / Stop Starts and stops computer data logging. When data
logging is started, logged data is placed in the active Excel worksheet
and the worksheet is protected to provide data security.
Logging Properties Allows the user specify computer data logging
parameters. The data logging sample interval (2 seconds to 999 hours),
total data logging period (2 seconds to 49 days) and computer data
log alarms may be specified through this menu. These parameters are
independent of the 451’s internal data logging functions.
Integrate Allows the user to select timed integrated dose
measurement or user controlled integrated dose measurement. The
user may also specify the integration period for timed integration from
1 minute to 999 hours. When timed integrate mode is selected, the
451 Assistant for Excel stops the integrated dose measurement after
the specified integration period has expired.
When timed integrate mode is not selected, the user controls the
integration period and the integration time is displayed in real time.
The integrated dose and average dose rate are recorded in the active
Excel worksheet when integration is complete.
Download Log The 451 is capable of internally logging data at a
user defined interval from 1 to 255 seconds with a total capacity of up
to 2700 data points. This feature allows the user to download the
451’s internal data log into the active Excel worksheet
Options Allows the user to change many properties of the
451 Assistant for Excel and several of the 451s’ properties. The user
may configure the various visual and audio alarm features of the
451 Assistant for Excel for each data point that is logged. The
451 Assistant can be configured to use 2 or 4 alarm states. The four
alarm states are acknowledged normal (below alarm level),
unacknow-ledged normal, acknowledged alarm and unacknowledged
alarm. The user may also select the communication port used by the
451 (default is COM1). The 451’s internal data logging parameters
and alarm settings may also be changed from the Options menu.
|
|