Monitor 4 and 4EC Analog Survey Meters |
| Detector Monitor 4: Halogen-quenched uncompensated GM tube with thin mica window 1.5-2.0 mg/cm2 thick. Monitor 4EC: Halogen-quenched GM tube, energy compensated sidewall 2mm tin filter. Thin mica window 1.5-2.0 mg/cm2 thick. Operating Range 0-.5, 0-5, 0-50 mR/hr and 0-500, 0-5,000, 0-50,000 CPM or 0-500 µSv/hr and 0-50 mR/hr. Range Switch X1, X10, X100, Battery Check. Accuracy ±10% of full scale (referenced to Cs-137) Monitor 4 Energy Sensitivity Detects alpha down to 2.5 MeV; typical detection efficiency at 3.6 MeV is greater than 80%. Detects beta at 50 keV with typical 35% detection efficiency. Detects beta at 150 keV with 75% typical detection efficiency. Detects gamma and x-rays down to 10 keV typical through the window, 40 keV minimum through the sidewall of the detector. Normal background is approximately 10-20 CPM. Monitor 4EC Energy Sensitivity Same as Monitor 4 except the energy response for gamma and x-rays through the detector sidewall is flat within +61% or -26% over the range of 40 keV to 100 keV, and within +35% or -17% over the range of 100 keV to 1.3 MeV (referenced to Cs-137). |
|