JRT Associates   Nuclear Medicine
Radiation Oncology
Emergency Preparedness
PET Imaging
Nuclear Cardiology
Medical Physics
Home Products Contest Central Newsletters JRT Calibrations Contact JRT
SPECIFICATIONS

Telepole Wide Range Gamma Survey Meter

RADIOLOGICAL CHARACTERISTICS
Remote Detector GM-tube ZP-1301 and ZP 1201
Measuring Range 0.05 mR/h to 1000 R/h (0.5 μSV/h to 10 Sv/h)
Accuracy ± 10 of reading, within the measuring range
Energy Response ± 20% at 70 keV to 2 MeV related to 137Cs
Sensitivity 0.3 cps/mR/h (low range), 17 cps/mR/h (high range)

ELECTRICAL CHARACTERISTICS
Power Supply Two 1.5 Volt C-type alkaline cells
Battery Life approximately 70 hours of continuous use (30% less if backlight is on continuously)

MECHANICAL CHARACTERISTICS
Meter dimensions (w x l x h) 93 x 148 x 56 cm (3.7 x 5.8 x 2.2 in)
Probe head length 226 cm (8.9 in)
Probe head diameter 33 cm (1.3 in)
Weight 2.55 Kg (7.29 lb)
Casing black anodized aluminium

ENVIRONMENTAL CHARACTERISTICS
Temperature Operation: -10° to 50°C (15 to 120° F ); Storage: -20° to 60°C (-5 to    140° F ); Relative humidity: 10 to 95% RH (non-condensing)