Model 14C Survey Meter |
| Compatible Detectors G-M, scintillation Meter Dial Typically 0 - 2 mR/hr and cpm, BAT TEST (others available ) Multipliers X0.1, X1, X10, X100, X1000 Linearity Reading within plus or minus 10% of true value with detector connected Connector Series "C" (others available ) Internal Detector Energy compensated G-M (used with X1000 scale only) Energy Response Within plus or minus 15% of true value between 60 keV - 3 MeV (internal detector only ) Threshold 30 mV plus or minus 10 mV Audio Built in unimorph speaker with ON/OFF switch (greater than 60 dB at 2 feet ) Calibration Controls Accessible from front of instrument (protective cover provided ) |
High Voltage 900 volts Threshold 30 mV plus or minus 10 mV Response Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading Reset Push-button to zero meter Power 2 each "D" cell batteries (housed in sealed compartment that is externally accessible ) Battery Life Typically greater than 1000 hrs.(battery condition can be checked on meter ) Meter 2.5" (6.4 cm) arc, 1 mA analog type Construction Cast and drawn aluminum with beige polyurethane enamel paint Temperature Range -4° F(-20° C) to 122° F(50° C) May be certified for operation from -40° F(-40° C) to 150° F(65° C) Size 6.5" (16.5 cm)H X 3.5" (8.9 cm)W X 8.5" (21.6 cm)L Weight 3.5 lbs (1.6 kg) including batteries |