JRT Associates   Nuclear Medicine
Radiation Oncology
Emergency Preparedness
PET Imaging
Nuclear Cardiology
Medical Physics
Home Products Contest Central Newsletters JRT Calibrations Contact JRT
SPECIFICATIONS

PM-1703 Gamma and Gamma/Neutron Radiation Pagers

GAMMA
Detector CsI(Tl) scintillator.
Measurement Range 0.1 to 50 μSv/h (10 - 5000 μR/h). Equivalent dose rate 137Cs.
Energy Range 0.06 to 3.0 MeV.

NEUTRON
Detector
Lil(Eu) scintillator.
Energy Range Thermal to 14 MeV.

GENERAL

Measurement Time 0.25 seconds.
Detection Sources Meets ITRAP (Illicit Trafficking Radiation Detection Assessment Program).
Alarm Types Audio tone and/or vibration.
Power Requirements One AA size battery. Life expectancy is approximately 800 hours.
PC Communications Through IR-interface.
Data Collection 900 data points stored in non-volatile memory.
Operating Temperature -30 to 50 °C (-22 to 122 °F).
Water Tightness Meets IP67.
Drop Test 0.7 m (27.5”) onto concrete surface.
Dimensions 87 x 72 x 32 mm (3.4” x 2.8” x 1.2”). Not including clip.
Weight PM1703M: 180 g (6.3 oz). Including battery. PM1703GN: 230 g (8.0 oz). Including battery.