JRT Associates   Nuclear Medicine
Radiation Oncology
Emergency Preparedness
PET Imaging
Nuclear Cardiology
Medical Physics
Home Products Contest Central Newsletters JRT Calibrations Contact JRT
SPECIFICATIONS

SOR/T AND SOR/R Personal Electronic Dosimeters

PHYSICAL CHARACTERISTICS
Hp(10) dose equivalent measurements
Flash Gamma Dose Measurement 5 cGy to 10 Gy (SOR/T)
Relative Error of Flash Measurement +/- 30 % over measurement range (SOR/T)
Ambient Gamma Dose Measurement Range 1μGy to 10 Gy
Gamma Dose Rate Measurement Range From background to 10 Gy/h
Gamma Dose Rate Display From 1 or 10 μGy/h to 10 Gy/h
Saturation Indication (above 10 Gy/h)
Relative Error of Ambient Measurement < +/- 20 % over the dose measurement range
Energy Response < +/- 20 % in the range 60 keV to 2 MeV;  < +/- 50 % in the range 2 MeV to 6 MeV

 

MECHANICAL CHARACTERISTICS
Dimensions 80.4 x 48 x 9 mm (flat housing) (3.16 x 1.85 x 0.35 in)
Weight 55 g (1.94 oz)

ENVIRONMENTAL CHARACTERISTICS

  • Temperature -20° to + 50°C (-4°F + 122°F) (normal operating range, standard battery 3V LiMno2 CR2430) -40° to + 50°C (-40°F + 122°F) (option with battery module 3.6V LiSoCl2)
  • TREE protection (SOR/T)
  • Resistant to EMP, EMC, radars
  • Resistant to water immersion, drops, shocks, vibrations, low pressure, initial conditions, NBC environmental conditions
  • Complies with the following standards: Meets MIL-STD-810 and MIL-STD-461 requirements; qualified by most of the NATO military laboratories; complies with IEC 1283, ANSI 42-20 and NATO D104