JRT Associates   Nuclear Medicine
Radiation Oncology
Emergency Preparedness
PET Imaging
Nuclear Cardiology
Medical Physics
Home Products Contest Central Newsletters JRT Calibrations Contact JRT
SPECIFICATIONS

DMC 2000XB Personal Electronic Dosimeter

Physical Characteristics
Complies with IEC 61526
Display Units mSv or mrem
Measurement Range  Dose: 1 μSv to 10 Sv (0.1 mrem to 1000 rem);  Dose Rate 0.1 μSv/h to 10 Sv/h (0.01 mrem/h to 1000 rem/h)
Display 0.01 mSv/h to 10 Sv/h (1 mrem/h to 1000 rem/h)
Linearity < +/- 10 % up to 1 Sv/h (100 rem/h); < +/- 25 % up to 10 Sv/h (1000 rem/h)    for X, gamma >60 keV and beta; < +/- 25 % up to 3 Sv/h (300 rem/h) for X < 60 keV
Energy Response X, gamma rays: 20 keV to 6 MeV; beta Emean > 60 keV (Emax: 0.22MeV to 2.3 MeV)
Accuracy < +/- 5 % (137Cs, 0.2 mSv/h)
  Mechanical Characteristics
Dimensions 84 x 48 x 17.5mm (3.3 x 1.9 x 0.7 in)
Weight with Battery < 70 g (2.2 oz)
Attached by replaceable clip


Electrical Characteristics

Li MnO2, CR2450 battery, battery life > 1 year (8 hrs/d) or 6 months continuous use

Environmental Characteristics

Temperature -10°C to +50°C (14°F to 122°F)
Humidity < 90 % at 42°C (108°F)
Storage -30°C to +71°C (-22°F to 160°F)
Shock, vibration and drop resistant
EMC complies with and exceeds standards