SPECIFICATIONS

identiFINDER™ Isotope Identifier

FEATURES
Functions Nuclide identification, spectrum analysis, dose rate calculation, total dose display, source finding

Integrated Electronics Multi-Channel-Analyzer, PMT preamplifier, spectroscopy amplifier, power supply

  • Standard Instruments
    • - identiFINDER-NG+ w/ 1.4"x 2" NaI & GM detectors
    • - identiFINDER-NGH+ w/ 1.4"x2" NaI, 3He & GM detectors
    • - identiFINDER-NG ultra w / 1.4”x2” NaI & GM detectors
    • - identiFINDER-NGH ultra w/ 1.4”x2” NaI, 3He & GM detectors
  • Specials 1.2"x1.4" LaBr with a resolution of <3.3% at 662 keV
  • Watertight (-U versions) Up to 1 atm (33’) emersion
  • Telescopic (-X versions) ~4’ to 8’ extension
  • Safeguards 1" x 1” Nal (TI) tungsten shielded detector and specialized SNM firmware

Stabilization In the non-Ultra models a built-in 137Cs reference source (<15 nCi/500 Bq) is used for online stabilization and in-situ calibration without user interaction. With either stabilization method this special identiFINDER feature allows operation over temperatures between -4 and 122 °F (-20 to
55 °C )

Software The identiFINDER provides for easy storage of up to 100 spectra and rapid transfer to a PC for reach-back and/or qualitative in-situ analysis with the software supplied

 

USER SELECTABLE NUCLIDE LIBRARY
There are 74 reference spectra of radionuclides stored in 6 libraries (Nuclear, Industrial, Medical, Customs, OSI, and Security). All sub-libraries except OSI can be edited by adding or deleting specific nuclides from the list. Ten (10) reference spectra can be measured by the user and added to the predefined library spectra. Identification is done by a template-matching correlation procedure

PHYSICAL DIMENSIONS
Weight 2.75lbs (1250g) with 1.4" x 2" Nal and batteries
Temperature Range -20 to 55 °C (-4 to 122 °F)
Protection water proof, dust tight
Protection class: IP 54
Drop Test 2' 8" on concrete
Dimensions 9.8” x 3.7” x 3”

SPECTROMETRY SYSTEM SPECIFICATIONS
HV-Bias 200 - 1275 V automatically set to suit individual detector
Shaping type Digital filter
INL, top 99%: <0.05%
DNL, top 99%: <0.1%
Spectrum length 1024 channels
Pileup Rejection <100 nS, pulse pair res.
Throughput Rate
>100,000 cps
Input Rate >350,000 cps
Spectrum Memory 100 spectra at 1024 channels
Real Time Presets 1 s - 1,000,000 s
Live Time Presets 1 s - 1,000,000 s

DOSE/DOSE MEASUREMENT SPECIFICATIONS
Sensitivity >10,000 cps/mrem for 1.4"x2" Nal (TI) detector
Dose-Rate Range 1 μrem/h - 100 rem/h
Dose Range 10 μrem - 100 rem
Energy Range Nal: 15keV - 3.0 MeV; GM: 60 keV - 1.6 MeV
Alarm Levels Gamma: Four preset levels
Neutron Separate neutron alarms for 3He model, with blue indicator lamp